Hot blast stove thermocouple calibration furnace parameter _ instrument

by:JVTIA     2020-10-29
( 1) Test temperature range 300 ~ 1200 ℃ ( 2) Furnace long p 600? ( 3) Thermocouple calibration furnace hot blast stove furnace high temperature point deviates from the furnace tube of no more than 20? ( 4) In the furnace temperature is very high + 20? The temperature gradient 0 or less. 4℃/? ( 5) The power supply voltage: 50 hz220v plus or minus 10% ( 6) Heat up a large power: 5 kva ( 7) Heating time: from room temperature to 800 ℃, the heating power is not more than 5 kva takes about 30 min ( 8) Dimension: high * length * width = 1200 * 700 * 400 ( 9) Weight: about 37? Need to know: (order 1) This type of hot blast stove thermocouple calibration furnace platinum rhodium 30 - cannot be used for testing standards Platinum rhodium thermocouple and Ⅱ level precision industrial platinum rhodium 30 - Platinum rhodium thermocouple 6, if you want to buy used to detect the platinum rhodium - 30 Platinum rhodium thermocouple calibration furnace, we can arrange production, the test temperature range of 1000 ~ 1550 ℃, all comply with corresponding code requirements. ( 2) If users have different requirements to the height of the furnace body or appearance size, can be clearly put forward in order ( Or attach a shape size chart) According to the request of users, our company supply.