Hot blast stove thermocouple calibration furnace parameter _ instrument
by:JVTIA
2020-10-29
(
1)
Test temperature range 300 ~ 1200 ℃ (
2)
Furnace long p 600?
(
3)
Thermocouple calibration furnace hot blast stove furnace high temperature point deviates from the furnace tube of no more than 20?
(
4)
In the furnace temperature is very high + 20?
The temperature gradient 0 or less.
4℃/?
(
5)
The power supply voltage: 50 hz220v plus or minus 10% (
6)
Heat up a large power: 5 kva (
7)
Heating time: from room temperature to 800 ℃, the heating power is not more than 5 kva takes about 30 min (
8)
Dimension: high * length * width = 1200 * 700 * 400 (
9)
Weight: about 37?
Need to know: (order
1)
This type of hot blast stove thermocouple calibration furnace platinum rhodium 30 - cannot be used for testing standards
Platinum rhodium thermocouple and Ⅱ level precision industrial platinum rhodium 30 -
Platinum rhodium thermocouple 6, if you want to buy used to detect the platinum rhodium - 30
Platinum rhodium thermocouple calibration furnace, we can arrange production, the test temperature range of 1000 ~ 1550 ℃, all comply with corresponding code requirements.
(
2)
If users have different requirements to the height of the furnace body or appearance size, can be clearly put forward in order (
Or attach a shape size chart)
According to the request of users, our company supply.