Easily ignored when thermocouple calibration problem

Verification personnel in the process of calibrating thermocouple, for terminal sand, thermocouple, common problems such as short circuit or strapping deviating from the geometrical center to the test data is not accurate, generally can be found easily handle in time, but is easy to overlook some questions affect the test results.

1. The length of the thermocouple

JJG351 - 1996 'working with inexpensive metal thermocouple' verification regulation specified in the thermocouple length not less than 750 mm, for the length of the thermocouple is given by, because considering the thermocouple temperature measurement area after you leave to have enough wide temperature gradient zone. Thermocouple thermoelectric emfs is produced in the region, to effectively prevent the hot end of thermocouple ( Measurement end) The quantity of heat to the cold end ( Terminal) , the most basic way is to have enough of the cold junction of the thermocouple distance away from the hot end. In general error caused by the length of the thermocouple is not enough is negative, the revised is positive. The shorter the length, the greater the error is, therefore, before the furnace charging test to determine the length of the thermocouple.

2. Thermocouple wire bending

thermocouple wire is fine and soft, easy to deformation, when accidentally silk such as folding, twisted plastic deformation, make the hot electrode pair wire produces stress changed the thermocouple thermoelectric properties, so that the deformation of the accuracy of the thermocouple measurements. Therefore, must keep the thermocouple wire straight before test.

3. Thermocouple wire contaminated

thermocouple wire contaminated, even by oxidation, can make the hot electrode surface gloss accidentally, dark black, then the hot electrode unstable thermoelectric properties, the accuracy of the measurement data is poorer, therefore, to clean electrodes have pollution, eliminate pollution layer.

4. The effects of response time

contact method of temperature measurement is the basic principle of temperature measuring element to reach thermal equilibrium with the object being measured. When measuring temperature, therefore, need to keep a certain time and to make both reach thermal equilibrium. Keep the length of time, with temperature measuring elements related to the thermal response time. And thermal response time depends on the structure of the sensor and measurement conditions, the difference is great. So, in the process of daily check according to the different types of thermocouple to choose the appropriate heating rate, thermal equilibrium time.

5. The influence of the insulation resistance

thermocouple under high temperature, the insulation resistance rapid decrease with the temperature increasing, thus will lead to leakage current, the current through insulation insulation resistance had fallen into the instrument, the instrument instability or producing measurement error. Before loading furnace thermocouple, therefore, don't ignore the insulation resistance test, only when meet the requirements of verification regulation, temperature tolerance test can proceed.

GET IN TOUCH WITH Us
recommended articles
NEWS
At present, more and more customers on the market are just beginning to attach great importance to the hidden dangers of the application of air disc machines and equipment.
Ventilation duct manufacturers talk about three welding methods for fabric ducts: Method 1: Manual welding.
It should be erected by the scaffolder according to the special safety construction organization design (construction plan) or safety technical measures.
Compliance Standards and Common Problems in Ventilation Duct Equipment Many bosses generally ignore the smoke exhaust system and natural ventilation planning of the restaurant kitchen when they open a new restaurant, which eventually leads to the closure of the hotel for renovation, renovation or sale of serious losses.
Ventilation pipe ventilation pipe manufacturer is a hollow pipe used for ventilation.
When we want to enter an industry and decide to dedicate our light and heat, our calm choice is to analyze this market well.
How to ensure the construction quality of dust removal fabric duct ventilation duct? In order to promote the construction of dust-removing fabric duct ventilation ducts, we should formulate some drastic measures.
How does the ventilation duct manufacturer ensure the construction quality of the ventilation duct for the dust removal fabric duct? 1.
There are common pipes everywhere in daily life, and it is not clear what to do.
Ventilation ducts are metal or composite ducts for ventilation in industrial and civil buildings.
no data

CONTACT DETAILS

Phone: +86 13816377866
Email: jimmy011@126.com
Add:NO999, Sunqiao RD, Pudong area, Shanghai, China
 

CONTACT US FOR FREE SAMPLE !

For more products related question ,Welcome to inquiry us .

Copyright © 2019 Shanghai Jiutian Automation Equipment Co., Ltd. | Sitemap
Customer service
detect